Force Sensor Calibration — 4 lb Spring Test
Digipot 159 Sterilization: heat to 55°C for 24 hrs 5-point cal Linear fit

All
None
Pre-steril measured
Pre 5-pt line
Pre 5-pt ideal pts
Pre 0+8.88 line
Pre 0+8.88 ideal pts
Post 0 & 4 lb measured
Post 0+4 line
Post 0+4 ideal pts
Post 0+8.9 line
Post 0+8.9 ideal pts
Post-steril measured
Post 5-pt line
Post 5-pt ideal pts
Projection error at ~49 lb vs actual measured
ModelProjected (mv)Actual measured (mv)Error (mv)Error (lb equiv.)
Pre 5-pt @ 49.44 lb2264.12236+28.1+1.1
Pre 0+8.88 @ 49.44 lb2040.32236−195.7−7.5
Post 5-pt @ 49.35 lb2381.62340+41.6+1.5
Post 0+4 @ 49.35 lb2474.02340+134.0 (+5.73%)+4.9
Post 0+8.9 @ 49.35 lb2479.02340+139.0 (+5.94%)+5.0
Ideal decade forces by model
Force (lb) Pre 5-pt Pre 0+8.88 Post 0+4 Post 0+8.9 Post 5-pt
All measured forces — measured vs estimated
Force (lb) Pre 5-pt Pre 0+8.88 Post 0+4 Post 0+8.9 Post 5-pt
Italic values are model estimates. Normal weight values are direct measurements. 0 and 4 lb pre/post values are means of repeated readings.